Please enable cookies, and supports 1024 × 768 resolution
  Home  
  About Linkjoin  
  Products  
  User Forum  
  Services  
  Contact Us  
welcome to linkjoin website
    Model MATS-2010H

     
  Model MATS-2010H Hysteresisgraph  
     
Automatic measurement of the demagnetization curve of permanent magnetic material such as ferrite, AlNiCo, NdFeB, SmCo, etc. Accurate measurement of the magnetic characteristic parameters of remanence Br, coercive force HcB, intrinsic coercive force HcJ and maximum magnetic energy product (BH)max.

Windows measurement software applied simply. The product conforms to China National Standards GB/T 3217-92, GB 11209-89 and international standard IEC 404-5.

Adopt ATS structure, users can customize different configuration as required: According to the intrinsic and size of measured sample  to decide the electromagnetic size and corresponding testing power supply; Select different measuring coil and probe according to the option of measuring method. Decide to whether choosing fixture in accordance with sample shape. Decide to whether selecting heating installation or not according to the test requirement.

      General Features       Software Features       Software Screen       Technical Data       Product Family  

System Specifications

Test AlNiCo standard sample, measuring error on entire magnetic hysteresis cycle estimated: B: 1%, H: 1%

Parameters measured Br (%) HcB (%) HcJ (%) (BH)max (%)
Uncertainty (k = 2) 1 1 1 2
Repeatability (constant temperature) ± 0.3 ± 0.3 ± 0.3 ± 1

Instrument Specifications

TPS-2000H Test Power Supply   DCT130 Electromagnet
    Output Voltage: 0 ~ ±100 V
    Output Current: ±20 A (maximum value)
    Stability: superior to 0.05%
    Output protection: overvoltage, overflowing, short-circuits protection

    Head Diameter: 130 mm
    Head Spacing: 0 ~ 50 mm (continual adjustable)
    Maximal Magnetic Field: 1.5 T (head spacing 10 mm, coil current 20 A)

TPS-2100H Test Power Supply DCT150 Electromagnet
    Supply Voltage: single phase, 220V 50Hz
    Output Voltage: 0 ~ ±120 V
    Output Current: ±30 A (maximum value)
    Output protection: overvoltage, overflowing, short-circuits protection
    Head Diameter: 150 mm shrink to 80 mm
    Head Spacing: 0 ~ 50 mm (continual adjustable)
    Maximal Magnetic Field: 2.4 T (head spacing 10 mm,coil current 30 A)
TPS-2200H Test Power Supply DCT200 Electromagnet
    Supply Voltage: single phase, 220V 50Hz
    Output Voltage: 0 ~ ±200 V
    Output Current: ±60 A (maximum value)
    Output protection: overvoltage, overflowing, short-circuits protection

    Head Diameter: 200 mm shrink to 80 mm
    Head Spacing: 0 ~ 50 mm (continual adjustable)
    Maximal Magnetic Field: 2.7 T (head spacing 10 mm, coil current 60 A)

TPS-2300H Test Power Supply DCT250 Electromagnet
    Supply Voltage: three phase, 380V 50Hz
    Output Voltage: 0 ~ ±300 V
    Output Current: ±60 A (maximum value)
    Output protection: overvoltage, overflowing, short-circuits protection

    Head Diameter: 250 mm shrink to 80 mm
    Head Spacing: 0 ~ 80 mm (continual adjustable)
    Maximal Magnetic Field: 2.8 T (head spacing 10 mm, coil current 60 A)

LZ-210B Fluxmeter   PC6023 A/D D/A card
    Range: 2.5, 5, 10, 20 mWb
    Full Range Precision: 0.5%
    Zero Shift: 2 μWb / 30 seconds (2.5 mWb)

    A/D Conversion: 12 bits, 4 channels
    Conversion Time: < 10 μs
    Conversion Accuracy: 0.1% (full range)
    D/A Conversion: 12 bits, 2 channels
    Conversion Accuracy: 0.1% (full range)

LZ-210H Magnetometer PC6016 A/D D/A card
    Range: 0.5, 1, 1.5, 3 T
    Full Range Precision: 0.5%
    Non-Linearity: ±0.2% (software correction)
    A/D Conversion: 16 bits, 8 channels differential input
    Conversion Time: < 10 μs
    Conversion Accuracy: 0.01% (full range)
    D/A Conversion: 16 bits, 2 channels
    Conversion Accuracy: 0.01% (full range)
Correlative Services
    Reference Hysteresis Graphs

        About Linkjoin | Site Map | Contact Us © 2000-2013 Linkjoin Technology Co, Ltd. All Rights Reserved.